TSURUGA MODEL 4160series (416B, 416C, 416D, 416K)
Quantity
DC/AC Voltmeter/, DC/AC Ammeter, Receiver
- ¡ö DIN size 72×36 mm Compact size series
- ¡ö Available with scaling function
- ¡ö Available with many output options
- Model Designation
-
416 ¡õ-¡õ - ¡õ - ¡õ ©¸¢Ù©¼ ¢Ú ¢Û ¢ÙInput Specification / Measuring Range
DC Input
¡¾ DC Voltage Measurement ¡¿ Code Measuring Range Input Resistance Accuracy *1 Overload D-02 ±99.99mV 100MΩ ±(0.05% of rdg + 3digit) DC±250V D-03 ±999.9mV 1MΩ ±(0.05% of rdg + 3digit) DC±250V D-04 ±9.999 V 1MΩ ±(0.05% of rdg + 3digit) DC±250V D-05 ±99.99 V 1MΩ ±(0.05% of rdg + 3digit) DC±250V ¡¾ DC Current Measurement ¡¿ Code Measuring Range Input Resistance Accuracy *1 Overload D-12 ±99.99μA 1KΩ ±(0.1% of rdg + 3digit) DC ± 10 mA D-13 ±999.9μA 100Ω ±(0.1% of rdg + 3digit) DC ± 50 mA D-14 ±9.999mA 10Ω ±(0.1% of rdg + 3digit) DC ±150 mA D-15 ±99.99mA 1Ω ±(0.1% of rdg + 3digit) DC ±500 mA D-16 ±999.9mA 0.1Ω ±(0.15% of rdg +5digit) DC ±3 A Receiver Input
Code Measuring Range Input Resistance Accuracy *1 Overload C-09 DC 1 to 5 V 1MΩ ±(0.1% of rdg + 3 digit) DC ±250 V C-19 DC 4 to 20 mA 25Ω ±(0.1% of rdg + 3digit) DC ±100 mA C-V1 DC 0 to 1 V 1MΩ ±(0.1% of rdg + 3digit) DC ±250 V C-V2 DC 0 to 5 V 1MΩ ±(0.1% of rdg + 3digit) DC ±250 V C-V3 DC 0 to 10 V 1MΩ ±(0.1% of rdg + 3digit) DC ±250 V C-A1 DC 0 to 1 mA 100Ω ±(0.1% of rdg + 3digit) DC ±50 mA AC Input (Average Value)
¡¾ AC Voltage Measurement ¡¿ Code Measuring Range Input Resistance Accuracy *1 Overload B-22 AC 99.99mV 1MΩ ±(0.1% of rdg + 10digit) AC 100 V B-23 AC 999.9mV 1MΩ ±(0.1% of rdg + 10digit) AC 100 V B-24 AC 9.999 V 1MΩ ±(0.1% of rdg + 10digit) AC 250 V B-25 AC 99.99 V 1MΩ ±(0.1 % of rdg +10digit) AC 250 V B-26 AC 699.9 V 10MΩ ±(0.2 % of rdg +10digit) AC 700 V ¡¾ AC Current Measurement ¡¿ Code Measuring Range Input Resistance Accuracy *1 Overload B-32 AC 99.99 µA 1KΩ ±(0.1% of rdg + 20digit) AC 10mA B-33 AC 999.9 μA 100Ω ±(0.1% of rdg + 20digit) AC 50mA B-34 AC 9.999 mA 10Ω ±(0.1 % of rdg +20digit) AC150mA B-35 AC 99.99 mA 1Ω ±(0.1 % of rdg +20digit) AC500mA B-36 AC 999.9 mA 0.1Ω ±(0.2 % of rdg +20digit) AC 3A B-37 AC 5.000 A 0.01Ω ±(0.2 % of rdg +20digit) AC 7A AC Input (Real Effective Value)
¡¾ AC Voltage Measurement ¡¿ Code Measuring Range Input Resistance Accuracy *3 Overload K-22 AC 99.99 mV 1MΩ ±(0.1% of rdg + 10digit) AC 100 V K-23 AC 999.9mV 1MΩ ±(0.1% of rdg + 10digit) AC 100 V K-24 AC 9.999 V 1MΩ ±(0.1% of rdg + 10digit) AC 250 V K-25 AC 99.99 V 1MΩ ±(0.1 % of rdg +10digit) AC 250 V K-26 AC 699.9 V 10MΩ ±(0.2 % of rdg +10digit) AC 700 V ¡¾ AC Current Measurement ¡¿ Code Measuring Range Input Resistance Accuracy *3 Overload K-32 AC 99.99μA 1KΩ ±(0.1% of rdg + 20digit) AC 10 mA K-33 AC 999.9μA 100Ω ±(0.1% of rdg + 20digit) AC 50 mA K-34 AC 9.999mA 10Ω ±(0.1 % of rdg +20digit) AC 150 mA K-35 AC 99.99mA 1Ω ±(0.1 % of rdg +20digit) AC 500 mA K-36 AC 999.9mA 0.1Ω ±(0.2 % of rdg +20digit) AC 3 A K-37 AC 5.000A 0.01Ω ±(0.2 % of rdg +20digit) AC 7 A ¢ÚSupply Voltage
Code Power Source Voltage 3 AC 100 V (90 to 132 V) 5 AC 200 V (180 to 264 V) B DC 9 to 32 V ¢ÛData Output
Code Specifications Output Impedance Allowable Load Resistance Blank For display only 03 Analog output DC 0 to 1 V Less than 0.1 Ω More than 1 kΩ 04 Analog output DC 0 to 5 V Less than 0.1 Ω More than 5 kΩ 05 Analog output DC 0 to 10 V Less than 0.1 Ω More than 10 kΩ 09 Analog output DC 1 to 5 V Less than 0.1 Ω More than 5 kΩ 23 Analog output DC 0 to 1 mA More than 5 MΩ 0 to 5 kΩ 29 Analog output DC 4 to 20 mA More than 5 MΩ 0 to 250 Ω BP BCD output (TTL level: positive logic£© BN BCD output (TTL level: negative logic£© DN BCD output £¨Transistor output source type£© CP High, low limit comparator output *1 Accuracy : Defined at 23 ± 5°C, 45 to 75£¥ RH *2 Temperature coefficient : Temperature application range 0 to 50°C *3 Accuracy : Defined at 23 ± 5°C, 45 to 75£¥ RH
However, it is only applicable to the more than 5% of sine wave input of the full scale.
(Crest factor: 4, In case of 416K–37 is 2)
(Peak value up to 1000 V for 416K-26)General Specifications
Display 0 – 9999 red LED (character height 15 mm), with zero-suppress function.
Over-range indication: Display flashes with 4 digit when exceeds 9999.
Decimal Point: Adjustment of optional setting is available from the panel front.Scaling function Refer the operation manual. Hold function Holds display value (BCD output and RS-483 output) Sampling rate Approximately 2.5 times/sec. Input configuration Single ended, floating input A/D conversion Dual slope integration system Noise rejection Normal mode (NMR): More than 50 dB (Except AC input)
Common mode (CMR): More than 110 dB (Except AC input)
Power source line mixing noise: 1000 VWithstand voltage Input terminals - COM of each output: AC 500 V for 1min.
Input / output terminals - Power source terminals: AC 1500 V for 1min. (*)
Power source terminals – External case: AC 1500 V for 1min.
(*): AC 1000 V in case of DC power source.Insulation resistance DC 500 V More than 100 MΩ Power source voltage AC power source voltage: AC90-132V or AC180-264V 50/60Hz
DC power source voltage: DC9-32VPower consumption Approx. 2 VA with AC 100 V
Approx. 170 mA with DC 12 V
Approx. 100 mA with DC 24 VOperating temperature 0 to 50°C Storage temperature -20 to 70°C Weight AC power source models Approx. 260 g
DC power source models Approx. 200 g