TSURUGA MODEL 3128
TSURUGA MODEL 3128 Panel Meter Digital/Printer Tsuruga Malaysia, Penang, Butterworth Supplier, Suppliers, Supply, Supplies | TECH IMPRO AUTOMATION SOLUTION SDN BHD
Quantity
DC Voltmeter/Ammeter, Receiver

  • ¡ö DIN 48×24 mm compact size.
  • ¡ö Available with offset and full scale -1999 to 1999 optional setting.
  • ¡ö Isolation between power supply and input.
  •  
Model Designation
3128 - ¡õ - ¡õ - ¡õ
    ¢Ù   ¢Ú   ¢Û

¢ÙMeasuring Input

Code Measuring Range Input Resistance Accuracy *1 Overload
02 ±199.9 mV 1 MΩ ±(0.1% of rdg + 1 digit) DC ±100 V
03 ±1.999 V 6 MΩ ±(0.1% of rdg + 1 digit) DC ±250 V
04 ±19.99 V 10 MΩ ±(0.2% of rdg + 1 digit) DC ±250 V
05 ±199.9 V 10 MΩ ±(0.2% of rdg + 1 digit) DC ±500 V
09 1 to 5 V 5 MΩ ±(0.2% of rdg + 2 digit) DC ±250 V
V0 0 to 100 mV 1 MΩ ±(0.2% of rdg + 2 digit) DC ±100 V
V1 0 to 1 V 1.3 MΩ ±(0.2% of rdg + 1 digit) DC ±250 V
V2 0 to 5 V 5 MΩ ±(0.2% of rdg + 1 digit) DC ±250 V
V3 0 to 10 V 10 MΩ ±(0.2% of rdg + 1 digit) DC ±250 V
12 ±199.9 µA 1 kΩ ±(0.2% of rdg + 1 digit) DC ±2 mA
13 ±1.999 mA 100 Ω ±(0.2% of rdg + 1 digit) DC ±50 mA
14 ±19.99 mA 10 Ω ±(0.2% of rdg + 1 digit) DC ±150 mA
15 ±199.9 mA 1 Ω ±(0.2% of rdg + 1 digit) DC ±500 mA
16 ±1.999 A 0.1 Ω *2 ±(0.3% of rdg + 1 digit) DC ±3 A
19 4 to 20 mA 10 Ω ±(0.2% of rdg + 2 digit) DC ±150 mA
A1 0 to 1 mA 200 Ω ±(0.2% of rdg + 1 digit£© DC ±30 mA
*1    Accuracy:
      Defined at 23 ± 5°C, 45 to 75% RH. 
      Temperature coefficient:
     ±300 ppm/°C (Temp. range: 0 to 50°C)
Calibrated by standard thermal power mV input in compliance with JIS C 1602.
      Standard contact compensation:
     Defined at ±2°C (Temp. range: 0 to 50°C)
Resistance Thermometer
Code Type of sensor Measuring Range Display Range Accuracy¡¡*2
12 Pt100Ω -199.9 to 199.9°C -199.9 to 199.9°C ±(0.2% of rdg + 0.3°C)
11 -199 to 600°C -199 to 650°C ±(0.4% of rdg + 1°C)
*1  Accuracy:
Defined at 23°C±5°C, 45 to 75% RH
     Coefficient of temperature:
    ±200 ppm/°C (Temp. Range 0 to 50°C)
     *2 Shunt resistance:
 External type

      ¢ÚDisplay Color

Code Description
R Red LED
£Ç Green LED

¢ÛOption

Code Input
Blank null
A01 100 digit off
A02 abusolute value display

General Specifications

Display 0 to 1999 red or green LED (character height: 10mm)
Zero suppress function (Selection of setting available)
Scaling function Full-scale display: -1999 to 1999
Offset display: -1999 to 1999
Offset fixing function Displace can be fixed to offset value when the input value is lower than the offset value. Selection of setting is available.
Decimal point Optional setting by the connector: (No isolation and measurement input)
Over-range display Condition of over display (setting selection ¢Ù or ¢Ú)
¡¡¢Ù Over display when exceeds 1999 or -1999
¡¡¢Ú Over display when the scale exceeds ±100% or 130%
Over display (setting selection ¢Û or ¢Ü)
¡¡¢Û 1¡õ¡õ¡õ or -1¡õ¡õ¡õ (¡õ is a blank)
¡¡¢Ü ¡õ or ¡õ (¡õ is a blank)
Hold function Holds measurement data (Not isolated from input)
Resolution 1/2000
Display cycle 100ms, 400ms, 1s, 2s, 4s, 5s Selection of setting available
Input configuration Single ended, floating input
A/D conversion ¨S-Σ conversion system
Noise rejection Normal mode £¨NMR) More than 40 dB
Insulation resistance DC 500 V More than 100 MΩ
Withstand voltage Between input terminals and external case: AC 1500 V for 1min.
Between power terminals and external case: AC 1500 V for 1min.
Between power supply terminals and input terminals: AC 500 V for 1min.
Power supply DC 4.75 to 26 V
Power consumption
  Red LED Green LED
DC 5 V 110 mA 150 mA
DC 12 V 45 mA 80 mA
DC 24 V 30 mA 50 mA
Operating temperature 0 to 50°C
Storage temperature -20 to 70°C
Weight Approx. 40 g

Please leave your enquiry here, we will reply as soon as possible.


*only support gif, jpeg, jpg, png, pdf
Switch To Desktop Version